Header #17

The Buzz Adams Microfabrication Facility Instrumentation


Major Equipment in the Buzz Adams Microfabrication Facility

Click for larger images
Photolithography Equipment
ABM Mask

ABM UV Flood and Mask Alignment System

The i-line near-UV flood source produces light with an intensity of 21 mW/cm2, with wavelengths below 350 nm being filtered out. Automatic exposure timing down to 0.1 s, stage adjustments (x, y, z, and Θ) with 1 μm precision.
spin coater

Brewer CEE100 Programmable Spin Coater & Thermolyne PMC Digital Hot Plate

The spin coater has ten programmable presets, each with timed and stepped control of acceleration rates (1-30000 r/m/s) and held angular speeds (1-6000 r/m), with repeatability of ±5 r/m. The microprocessor controlled PMC hot plate holds timed and stepped temperatures from 38 to 370° C (± 2° C).
Diagnostic  Scope

Diagnostic and Circuit-testing Microscopes

One of three inspection microscopes in our facility, this Zeiss microscope features 5x, 10x, 20x, 50x and 100x objective lenses, and dark and bright field illumination. An attached camera enables video and still image capturing.
SPI Plasma Prep

Diagnostic and Circuit-testing Microscopes

One of three inspection microscopes in our facility, this Zeiss microscope features 5x, 10x, 20x, 50x and 100x objective lenses, and dark and bright field illumination. An attached camera enables video and still image capturing.
Plasma Etching
Plasmalab100

Oxford Plamalab System 100

This instrument is used for plasma etching of high surface ratio features
HF etching station

Hydroflouric Acid Etching Station

Set in a dedicated fume hood with a deep basin, this etching station is used to wet etch glass microfluidic devices using a buffered oxide etch process.
Film Deposition
Lesker Deposition Chamber

Lesker thin film deposition system

This instrument deposits films—primarily metals— using one of two DC magnetron sputtering sources in a low temperature environment. An Inficon quartz crystal monitors deposition rate and thickness.
Plasmalab80

Oxford Plasmalab 80 Plus PECVD

The new plasma-enhanced deposition system uses NH3, N2O, N2, CF4, and SiH4/O2 to deposit siliconbased, nitride and oxide films on various substrates for superconductor fabrication at low temperatures.
Semiconductor Characterization
Ellipsometer

HORIBA Jovin Yvon UVISEL - Spectroscopic Ellipsometer

The UVISEL ellipsometer measures features down to 250 nm, with a spectral range from 250 to 2100 nm—ideal for measuring semiconductor features and thin film deposition.
Evaluation
Tencor Profilometer

Tencor Alphastep 200 Profilometer

The Alphastep profilometer is used to measure features down to low micrometer heights, with scan lengths from 80 to 10000 microns, and resolution down to 5 nanometers.
Miscellaneous
furnace

Fisher Programmable Muffle Furnace

Used primarily for glass bonding, this furnace allows for timed and stepped control of temperatures from ambient to 1125° C.
labconco hood

Labconco Vertical Clean Bench

This work station is used for mixing and degassing substrate monomers, such as PDMS, prior to molding and curing. Air handling provides an ISO Class 5 (US FED Class 100) work environment.
Microchip Machine Shop
laser engraver

Universal CO2 Laser Engraving System

In addition to myriad other uses, this equipment is used to ablate microchannels in polymer plastics (such as PDMS and PMMA) and glass, with feature sizes down to ~100 μm across.
heatable press

Carver heatable lab press